The short answer. A UV meter can show how much peptide is in a sample. At 280 nm, only a few amino acids take up much light. The main ones are tryptophan and tyrosine. A peptide with neither barely shows up. Seven of the eight peptides in our table have neither, so the test does not work for them.

How the reading works

Shine UV light through a sample, and some of it is absorbed. The amount absorbed, called absorbance, rises with how much peptide is there. To turn absorbance into a concentration, you need one number for the peptide: how strongly each molecule absorbs. That number is called the molar absorption coefficient, or ε (epsilon).

Absorbance = ε × concentration (mol/L) × path length (cm). Most cuvettes have a 1 cm path.

Predicting ε at 280 nm

For proteins, Pace and colleagues found that ε at 280 nm can be predicted from three building blocks:1

  • Each tryptophan (W) adds 5,500.
  • Each tyrosine (Y) adds 1,490.
  • Each cystine (a pair of linked cysteines) adds 125.

The same paper notes two limits. The prediction is reliable when tryptophan is present and less reliable without it. And measuring ε is better than predicting it.1 The formula was built on proteins, so for short peptides treat it as an estimate.

Our peptides at 280 nm

Peptide W Y Predicted ε at 280 nm (M⁻¹cm⁻¹)
Kisspeptin-10 1 1 6,990
Selank 0 0 None
Semax 0 0 None
KPV 0 0 None
Epitalon 0 0 None
BPC-157 0 0 None
Thymosin Alpha-1 0 0 None
ARA-290 0 0 None

Counts from the PubChem sequences. None of these sequences has a cysteine.3

Only Kisspeptin-10 has the building blocks this method needs. For the other seven, a reading at 280 nm would be close to zero, and the formula gives no ε to convert it with.

Ipamorelin is a special case. It contains naphthylalanine, an unnatural building block the formula does not cover. Its ε at 280 nm would have to be measured, not predicted.

A worked example

Take Kisspeptin-10 at 0.1 mg/mL. Its molecular weight is 1,302.44, so that is about 77 micromolar. With ε = 6,990 and a 1 cm path, the predicted absorbance is about 0.54. If a sample reads very differently from what you expect, check the calculation and the blank first. Remember too that the weighed powder includes water and a salt partner, so 0.1 mg of powder holds less than 0.1 mg of peptide. See HPLC purity vs peptide content.

When 280 nm does not work

Every peptide bond absorbs light at 214 nm, so peptides without tryptophan or tyrosine can still be measured there.2 The trade-off is that many other substances absorb at that low wavelength too, so the solution must be clean and the blank matched. Other options include amino acid analysis, which measures the amino acids directly after breaking the peptide down.

The same limit applies to a lab report. A net content figure in milligrams cannot be checked with a 280 nm reading for most of the peptides above.

Sources

  1. Pace CN, Vajdos F, Fee L, Grimsley G, Gray T. How to measure and predict the molar absorption coefficient of a protein. Protein Sci. 1995;4(11):2411-23. doi:10.1002/pro.5560041120. PMID 8563639.
  2. Kuipers BJ, Gruppen H. Prediction of molar extinction coefficients of proteins and peptides using UV absorption of the constituent amino acids at 214 nm. J Agric Food Chem. 2007;55(14):5445-51. PMID 17539659.
  3. Sequences from PubChem: Kisspeptin-10, Selank, Semax, KPV, Epitalon, BPC-157, Thymosin Alpha-1, ARA-290, Ipamorelin.